Abstract
A new fiber optic microsensor has been developed for pressure measurement. The pressure microsensor is extremely small (360μm). Performance characteristics of the microsensor were evaluated over a pressure range from 0 to 250mmHg. Five sensors were compared to a National Institute of Standards and Technology traceable reference pressure. The microsensor had an accuracy, repeatability, and linearity better than 2% full-scale output (FSO) and a hysteresis of 4.5% FSO.
Original language | English (US) |
---|---|
Pages (from-to) | 283-287 |
Number of pages | 5 |
Journal | Journal of Biomechanics |
Volume | 36 |
Issue number | 2 |
DOIs | |
State | Published - Feb 1 2003 |
Keywords
- Extrinsic Fabry-Perot interferometer
- Fiber optic sensor
- Pressure
ASJC Scopus subject areas
- Biophysics
- Orthopedics and Sports Medicine
- Biomedical Engineering
- Rehabilitation