T1 - Noise performance of low-dose CT
T2 - Comparison between an energy integrating detector and a photon counting detector using a whole-body research photon counting CT scanner
AU - Yu, Zhicong
AU - Leng, Shuai
AU - Kappler, Steffen
AU - Hahn, Katharina
AU - Li, Zhoubo
AU - Halaweish, Ahmed F.
AU - Henning, Andre
AU - McCollough, Cynthia H.
N1 - Funding Information:
The project described was supported by Grant Nos. R01 EB016966 and C06 RR018898 from the National Institutes of Health in collaboration with Siemens Healthcare.
Publisher Copyright:
© 2016 Society of Photo-Optical Instrumentation Engineers (SPIE).
PY - 2016/10/1
Y1 - 2016/10/1
N2 - Photon counting detector (PCD)-based computed tomography (CT) is an emerging imaging technique. Compared to conventional energy integrating detector (EID)-based CT, PCD-CT is able to exclude electronic noise that may severely impair image quality at low photon counts. This work focused on comparing the noise performance at low doses between the PCD and EID subsystems of a whole-body research PCD-CT scanner, both qualitatively and quantitatively. An anthropomorphic thorax phantom was scanned, and images of the shoulder portion were reconstructed. The images were visually and quantitatively compared between the two subsystems in terms of streak artifacts, an indicator of the impact of electronic noise. Furthermore, a torso-shaped water phantom was scanned using a range of tube currents. The product of the noise and the square root of the tube current was calculated, normalized, and compared between the EID and PCD subsystems. Visual assessment of the thorax phantom showed that electronic noise had a noticeably stronger degrading impact in the EID images than in the PCD images. The quantitative results indicated that in low-dose situations, electronic noise had a noticeable impact (up to a 5.8% increase in magnitude relative to quantum noise) on the EID images, but negligible impact on the PCD images.
AB - Photon counting detector (PCD)-based computed tomography (CT) is an emerging imaging technique. Compared to conventional energy integrating detector (EID)-based CT, PCD-CT is able to exclude electronic noise that may severely impair image quality at low photon counts. This work focused on comparing the noise performance at low doses between the PCD and EID subsystems of a whole-body research PCD-CT scanner, both qualitatively and quantitatively. An anthropomorphic thorax phantom was scanned, and images of the shoulder portion were reconstructed. The images were visually and quantitatively compared between the two subsystems in terms of streak artifacts, an indicator of the impact of electronic noise. Furthermore, a torso-shaped water phantom was scanned using a range of tube currents. The product of the noise and the square root of the tube current was calculated, normalized, and compared between the EID and PCD subsystems. Visual assessment of the thorax phantom showed that electronic noise had a noticeably stronger degrading impact in the EID images than in the PCD images. The quantitative results indicated that in low-dose situations, electronic noise had a noticeable impact (up to a 5.8% increase in magnitude relative to quantum noise) on the EID images, but negligible impact on the PCD images.
KW - computed tomography
KW - electronic noise
KW - low dose
KW - photon counting
UR - http://www.scopus.com/inward/record.url?scp=85007227136&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85007227136&partnerID=8YFLogxK
U2 - 10.1117/1.JMI.3.4.043503
DO - 10.1117/1.JMI.3.4.043503
M3 - Article
AN - SCOPUS:85007227136
SN - 2329-4302
VL - 3
JO - Journal of Medical Imaging
JF - Journal of Medical Imaging
IS - 4
M1 - 043503
ER -