Abstract
This study reviewed 52 revision hip arthroplasties in which calcar cortical deficiencies were addressed by implantation of a cemented, modern- design head and neck replacement femoral prosthesis. Mean clinical follow-up was 38 months. Radiographic follow-up beyond 22 months was available for 34 patients (mean: 35.8 months). Mean Mayo clinical hip score (maximum: 80) improved from 28.2 preoperatively to 55.2 at last follow-up. No femoral component was revised for aseptic loosening. However, radiographic analysis demonstrated one prosthesis was probably loose and two were possibly loose. A 92.7% Kaplan-Meier 5-year implant survival rate free from radiographic loosening was projected. This type of prosthesis has yielded good clinical results and excellent early survival.
Original language | English (US) |
---|---|
Pages (from-to) | 313-318 |
Number of pages | 6 |
Journal | Orthopedics |
Volume | 22 |
Issue number | 3 |
State | Published - Apr 7 1999 |
ASJC Scopus subject areas
- Surgery
- Orthopedics and Sports Medicine