Abstract
A fast shear compounding method was developed in this study using only one shear wave push-detect cycle, such that the shear wave imaging frame rate is preserved and motion artifacts are minimized. The proposed method is composed of the following steps: 1. Applying a comb-push to produce multiple differently angled shear waves at different spatial locations simultaneously; 2. Decomposing the complex shear wave field into individual shear wave fields with differently oriented shear waves using a multi-directional filter; 3. Using a robust 2-D shear wave speed calculation to reconstruct 2-D shear elasticity maps from each filter direction; and 4. Compounding these 2-D maps from different directions into a final map. An inclusion phantom study showed that the fast shear compounding method could achieve comparable performance to conventional shear compounding without sacrificing the imaging frame rate. A multi-inclusion phantom experiment showed that the fast shear compounding method could provide a full field-of-view, 2-D and compounded shear elasticity map with three types of inclusions clearly resolved and stiffness measurements showing excellent agreement to the nominal values.
Original language | English (US) |
---|---|
Pages (from-to) | 1343-1355 |
Number of pages | 13 |
Journal | Ultrasound in Medicine and Biology |
Volume | 40 |
Issue number | 6 |
DOIs | |
State | Published - Jun 2014 |
Keywords
- 2-D shear wave speed
- Acoustic radiation force
- Comb-push
- Directional filter
- Shear compounding
- Shear wave elastography
ASJC Scopus subject areas
- Biophysics
- Radiological and Ultrasound Technology
- Acoustics and Ultrasonics