TY - JOUR
T1 - Correcting the dynamic response of a commercial esophageal balloon-catheter
AU - Cross, Troy J.
AU - Beck, Kenneth C.
AU - Johnson, Bruce D.
N1 - Funding Information:
T.J. Cross was supported by a joint Mayo Clinic and Griffith University postdoctoral fellowship. This study was supported by National Heart, Lung, and Blood Institute Grant HL-71478.
Publisher Copyright:
Copyright © 2016 the American Physiological Society.
PY - 2016/8
Y1 - 2016/8
N2 - It is generally recommended that an esophageal balloon-catheter possess an adequate frequency response up to 15 Hz, such that parameters of respiratory mechanics may be quantified with precision. In our experience, however, we have observed that some commercially available systems do not display an ideal frequency response (8-10 Hz). We therefore investigated whether the poor frequency response of a commercially available esophageal catheter may be adequately compensated using two numerical techniques: 1) an exponential model correction, and 2) Wiener deconvolution. These two numerical techniques were performed on a commercial balloon-catheter interfaced with 0, 1, and 2 lengths of extension tubing (90 cm each), referred to as configurations L0, L90, and L180, respectively. The frequency response of the balloon-catheter in these configurations was assessed by empirical transfer function analysis, and its working range was defined as the frequency beyond which more than 5% amplitude and/or phase distortion was observed. The working frequency range of the uncorrected balloon-catheter extended up to only 10 Hz for L0, and progressively worsened with additional tubing length (L90 3 Hz, L180 2 Hz). Although both numerical methods of correction adequately enhanced the working frequency range of the balloon-catheter to beyond 25 Hz for all length configurations (L0, L90, and L180), Wiener deconvolution consistently provided more accurate corrections. Our data indicate that Wiener deconvolution provides a superior correction of the balloon-catheter's dynamic response, and is relatively more robust to extensions in catheter tube length compared with the exponential correction method.
AB - It is generally recommended that an esophageal balloon-catheter possess an adequate frequency response up to 15 Hz, such that parameters of respiratory mechanics may be quantified with precision. In our experience, however, we have observed that some commercially available systems do not display an ideal frequency response (8-10 Hz). We therefore investigated whether the poor frequency response of a commercially available esophageal catheter may be adequately compensated using two numerical techniques: 1) an exponential model correction, and 2) Wiener deconvolution. These two numerical techniques were performed on a commercial balloon-catheter interfaced with 0, 1, and 2 lengths of extension tubing (90 cm each), referred to as configurations L0, L90, and L180, respectively. The frequency response of the balloon-catheter in these configurations was assessed by empirical transfer function analysis, and its working range was defined as the frequency beyond which more than 5% amplitude and/or phase distortion was observed. The working frequency range of the uncorrected balloon-catheter extended up to only 10 Hz for L0, and progressively worsened with additional tubing length (L90 3 Hz, L180 2 Hz). Although both numerical methods of correction adequately enhanced the working frequency range of the balloon-catheter to beyond 25 Hz for all length configurations (L0, L90, and L180), Wiener deconvolution consistently provided more accurate corrections. Our data indicate that Wiener deconvolution provides a superior correction of the balloon-catheter's dynamic response, and is relatively more robust to extensions in catheter tube length compared with the exponential correction method.
KW - Balloon-catheter
KW - Compensation
KW - Esophageal pressure
KW - Frequency response
KW - Wiener deconvolution
UR - http://www.scopus.com/inward/record.url?scp=84984660752&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84984660752&partnerID=8YFLogxK
U2 - 10.1152/japplphysiol.00155.2016
DO - 10.1152/japplphysiol.00155.2016
M3 - Article
C2 - 27402558
AN - SCOPUS:84984660752
SN - 8750-7587
VL - 121
SP - 503
EP - 511
JO - Journal of applied physiology
JF - Journal of applied physiology
IS - 2
ER -