TY - JOUR
T1 - Comparative Survival of Contemporary Cementless Acetabular Components Following Revision Total Hip Arthroplasty
AU - Wilson, Jacob M.
AU - Maradit-Kremers, Hilal
AU - Abdel, Matthew P.
AU - Berry, Daniel J.
AU - Mabry, Tad M.
AU - Pagnano, Mark W.
AU - Perry, Kevin I.
AU - Sierra, Rafael J.
AU - Taunton, Michael J.
AU - Trousdale, Robert T.
AU - Lewallen, David G.
N1 - Publisher Copyright:
© 2023 Elsevier Inc.
PY - 2023/7
Y1 - 2023/7
N2 - Background: The advent of highly porous ingrowth surfaces and highly crosslinked polyethylene has been expected to improve implant survivorship in revision total hip arthroplasty. Therefore, we sought to evaluate the survival of several contemporary acetabular designs following revision total hip arthroplasty. Methods: Acetabular revisions performed from 2000 to 2019 were identified from our institutional total joint registry. We studied 3,348 revision hips, implanted with 1 of 7 cementless acetabular designs. These were paired with highly crosslinked polyethylene or dual-mobility liners. A historical series of 258 Harris-Galante-1 components, paired with conventional polyethylene, was used as reference. Survivorship analyses were performed. For the 2,976 hips with minimum 2-year follow-up, the median follow-up was 8 years (range, 2 to 35 years). Results: Contemporary components with adequate follow-up had survivorship free of acetabular rerevision of ≥95% at 10-year follow-up. Relative to Harris-Galante-1 components, 10-year survivorship free of all-cause acetabular cup rerevision was significantly higher in Zimmer Trabecular Metarevision (hazard ratio (HR) 0.3, 95% confidence interval (CI) 0.2-0.45), Zimmer Trabecular MetaModular (HR 0.34, 95% CI 0.13-0.89), Zimmer Trilogy (HR 0.4, 95% CI 0.24-0.69), DePuy Pinnacle Porocoat (HR 0.24, 95% CI 0.11-0.51), and Stryker Tritanium revision (HR 0.46, 95% CI 0.24-0.91) shells. Among contemporary components, there were only 23 rerevisions for acetabular aseptic loosening and no rerevisions for polyethylene wear. Conclusion: Contemporary acetabular ingrowth and bearing surfaces were associated with no rerevisions for wear and aseptic loosening was uncommon, particularly with highly porous designs. Therefore, it appears that contemporary revision acetabular components have dramatically improved upon historical results at available follow-up.
AB - Background: The advent of highly porous ingrowth surfaces and highly crosslinked polyethylene has been expected to improve implant survivorship in revision total hip arthroplasty. Therefore, we sought to evaluate the survival of several contemporary acetabular designs following revision total hip arthroplasty. Methods: Acetabular revisions performed from 2000 to 2019 were identified from our institutional total joint registry. We studied 3,348 revision hips, implanted with 1 of 7 cementless acetabular designs. These were paired with highly crosslinked polyethylene or dual-mobility liners. A historical series of 258 Harris-Galante-1 components, paired with conventional polyethylene, was used as reference. Survivorship analyses were performed. For the 2,976 hips with minimum 2-year follow-up, the median follow-up was 8 years (range, 2 to 35 years). Results: Contemporary components with adequate follow-up had survivorship free of acetabular rerevision of ≥95% at 10-year follow-up. Relative to Harris-Galante-1 components, 10-year survivorship free of all-cause acetabular cup rerevision was significantly higher in Zimmer Trabecular Metarevision (hazard ratio (HR) 0.3, 95% confidence interval (CI) 0.2-0.45), Zimmer Trabecular MetaModular (HR 0.34, 95% CI 0.13-0.89), Zimmer Trilogy (HR 0.4, 95% CI 0.24-0.69), DePuy Pinnacle Porocoat (HR 0.24, 95% CI 0.11-0.51), and Stryker Tritanium revision (HR 0.46, 95% CI 0.24-0.91) shells. Among contemporary components, there were only 23 rerevisions for acetabular aseptic loosening and no rerevisions for polyethylene wear. Conclusion: Contemporary acetabular ingrowth and bearing surfaces were associated with no rerevisions for wear and aseptic loosening was uncommon, particularly with highly porous designs. Therefore, it appears that contemporary revision acetabular components have dramatically improved upon historical results at available follow-up.
KW - acetabular revision
KW - highly crosslinked polyethylene
KW - porous
KW - revision total hip
KW - uncemented
UR - http://www.scopus.com/inward/record.url?scp=85158104957&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85158104957&partnerID=8YFLogxK
U2 - 10.1016/j.arth.2023.03.093
DO - 10.1016/j.arth.2023.03.093
M3 - Article
C2 - 37028772
AN - SCOPUS:85158104957
SN - 0883-5403
VL - 38
SP - S194-S200
JO - Journal of Arthroplasty
JF - Journal of Arthroplasty
IS - 7
ER -