Automated Microfluidic System with Active Mixing Enables Rapid Analysis of Biomarkers in 5 μL of Whole Blood

Alan M. Gonzalez-Suarez, Gulnaz Stybayeva, William A. Carey, Alexander Revzin

Research output: Contribution to journalArticlepeer-review

Abstract

We developed a microfluidic device for the rapid analysis of biomarkers in small volumes of whole blood. This device includes an onboard plasma separation module connected to a downstream bioanalysis module in which plasma mixes with reagents and the results of a colorimetric assay are recorded. Actuation of onboard microvalves within a bioanalysis module creates active mixing conditions that allowed us to achieve solution homogeneity within 5 min. To demonstrate utility, we carried out glucose detection in our device. With 5 μL of whole blood as an input, our microfluidic device enabled a time-to-answer of 10 min with a limit of detection of 0.21 ± 0.04 mM for glucose. This device has immediate applications for rapid and sensitive monitoring of hypoglycemia at the point of care (POC). Furthermore, our automated microfluidic device represents a platform technology that may be used to detect other biomarkers in whole blood.

Original languageEnglish (US)
Pages (from-to)9706-9714
Number of pages9
JournalAnalytical Chemistry
Volume94
Issue number27
DOIs
StatePublished - Jul 12 2022

ASJC Scopus subject areas

  • Analytical Chemistry

Fingerprint

Dive into the research topics of 'Automated Microfluidic System with Active Mixing Enables Rapid Analysis of Biomarkers in 5 μL of Whole Blood'. Together they form a unique fingerprint.