Abstract
Objective. Attenuation of radio frequency (RF) signals limits the use of contrast echocardiography. The harmonic-to-fundamental ratio (HFR) of the RF signals compensates for attenuation. We tested whether HFR analysis measures the left ventricular nonperfused area under simulated experimental attenuation. Methods. Radio frequency image data from short axis systolic projections were obtained from 11 open-chest dogs with left anterior descending or left circumflex coronary artery occlusion followed by left atrial bolus injection of a perflutren microbubble contrast agent. Clinical attenuation was simulated by calibrated silicone pads interposed between the epicardial surface and the transducer to induce mild (7-dB) and severe (14-dB) reduction of the backscattered RF signals. Harmonic-to-fundamental ratio values were calculated for each image pixel for 0-, 7-, and 14-dB attenuation conditions and reproducibly showed a "perfused area" and a "nonperfused area." A reference nonperfused area was obtained by manual delineation in high-quality contrast scans. Results. Correlations of the HFR-detected and manually outlined perfusion defect areas were R = 0.92 for 0 dB, R = 0.94 for 7 dB, and R = 0.90 for 14 dB; the mean difference was less than 0.36 cm 2 (negligible) in all 3 attenuation settings. Conclusions. Attenuation compensation by our HFR method allows precise measurement of myocardial perfusion defect areas in contrast scans with simulated high level of attenuation.
Original language | English (US) |
---|---|
Pages (from-to) | 739-748 |
Number of pages | 10 |
Journal | Journal of Ultrasound in Medicine |
Volume | 26 |
Issue number | 6 |
DOIs | |
State | Published - Jun 2007 |
Keywords
- Contrast echocardiography
- Harmonic ultrasound
- Myocardial perfusion
- Radio frequency signal
- Ultrasound attenuation
ASJC Scopus subject areas
- Radiological and Ultrasound Technology
- Radiology Nuclear Medicine and imaging