Multi-Center, Multi-Vendor Validation of Simultaneous MRI-Based Proton Density Fat Fraction and R2* Mapping Using a Combined Proton Density Fat Fraction-R2* Phantom

  • Jitka Starekova
  • , David Rutkowski
  • , Won C. Bae
  • , Hung Do
  • , Ananth J. Madhuranthakam
  • , Vadim Malis
  • , Sheng Qing Lin
  • , Suraj Serai
  • , Takeshi Yokoo
  • , Scott B. Reeder
  • , Jean H. Brittain
  • , Diego Hernando

Research output: Contribution to journal › Article › peer-review

Abstract

Background: Fat and iron deposition confound measurements of R2* and proton density fat fraction (PDFF), respectively, yet their combined impact on reproducibility is poorly understood. Purpose: To evaluate the multi-center, multi-vendor reproducibility of PDFF and R2* quantification using a PDFF-R2* phantom. Study Type: Prospective multi-center, phantom study. Phantom: Commercial PDFF-R2* phantom with simultaneously controlled combination of PDFF (0%–30%) and R2* (50–600 s−1) values. Field Strength/Sequence: 1.5-T and 3-T, three-dimensional (3D) multi-echo, spoiled-gradient-echo sequences, in four different centers, each with a different vendor. Assessment: Two acquisition protocols were used, optimized for moderate R2* (Protocol 1) and high R2* (Protocol 2), respectively. The phantom was imaged multiple times at one of the centers to assess its stability. Statistical Tests: Intraclass correlation coefficient (ICC), linear regression analysis, reproducibility coefficient (RDC) and repeatability coefficient (RC). Results: Excellent agreement was observed for PDFF measurements between centers, vendors, field strengths, and protocols (ICC = 0.97). Stratified by protocol, excellent agreement was observed, with ICC = 0.96 (RDC = 6.2%) for Protocol 1 and ICC = 0.99 (RDC = 3.8%) for Protocol 2. Increased variability in PDFF measurements was observed with increasing PDFF and especially with higher R2*. Excellent agreement was observed for R2* between centers, vendors, field strengths, and protocols (ICC = 0.99). Stratified by protocol, strong agreement was observed, with ICC = 0.988 (RDC = 66.7 s−1) for Protocol 1 and ICC = 0.99 (RDC = 57.7 s−1) for Protocol 2. Higher variability in R2* measurements was observed in vials with higher PDFF or R2*. Stability tests demonstrated an ICC = 1.0 for PDFF and R2*, and RC of 0.4% for PDFF and 12 s−1 for R2*. Data Conclusion: Excellent PDFF and R2* reproducibility was observed across centers, vendors, field strengths, and acquisition protocols. Reproducibility decreased slightly with increasing PDFF and R2*, especially for PDFF measurements in vials with high R2*. Evidence Level: N/A. Technical Efficacy: Stage 1.

Original languageEnglish (US)
Pages (from-to)800-811
Number of pages12
JournalJournal of Magnetic Resonance Imaging
Volume62
Issue number3
DOIs
StatePublished - Sep 2025

Keywords

  • CSE-MRI
  • PDFF
  • R2*
  • phantom
  • quantification
  • reproducibility