Abstract
Background: Fat and iron deposition confound measurements of R2* and proton density fat fraction (PDFF), respectively, yet their combined impact on reproducibility is poorly understood. Purpose: To evaluate the multi-center, multi-vendor reproducibility of PDFF and R2* quantification using a PDFF-R2* phantom. Study Type: Prospective multi-center, phantom study. Phantom: Commercial PDFF-R2* phantom with simultaneously controlled combination of PDFF (0%–30%) and R2* (50–600 s−1) values. Field Strength/Sequence: 1.5-T and 3-T, three-dimensional (3D) multi-echo, spoiled-gradient-echo sequences, in four different centers, each with a different vendor. Assessment: Two acquisition protocols were used, optimized for moderate R2* (Protocol 1) and high R2* (Protocol 2), respectively. The phantom was imaged multiple times at one of the centers to assess its stability. Statistical Tests: Intraclass correlation coefficient (ICC), linear regression analysis, reproducibility coefficient (RDC) and repeatability coefficient (RC). Results: Excellent agreement was observed for PDFF measurements between centers, vendors, field strengths, and protocols (ICC = 0.97). Stratified by protocol, excellent agreement was observed, with ICC = 0.96 (RDC = 6.2%) for Protocol 1 and ICC = 0.99 (RDC = 3.8%) for Protocol 2. Increased variability in PDFF measurements was observed with increasing PDFF and especially with higher R2*. Excellent agreement was observed for R2* between centers, vendors, field strengths, and protocols (ICC = 0.99). Stratified by protocol, strong agreement was observed, with ICC = 0.988 (RDC = 66.7 s−1) for Protocol 1 and ICC = 0.99 (RDC = 57.7 s−1) for Protocol 2. Higher variability in R2* measurements was observed in vials with higher PDFF or R2*. Stability tests demonstrated an ICC = 1.0 for PDFF and R2*, and RC of 0.4% for PDFF and 12 s−1 for R2*. Data Conclusion: Excellent PDFF and R2* reproducibility was observed across centers, vendors, field strengths, and acquisition protocols. Reproducibility decreased slightly with increasing PDFF and R2*, especially for PDFF measurements in vials with high R2*. Evidence Level: N/A. Technical Efficacy: Stage 1.
| Original language | English (US) |
|---|---|
| Pages (from-to) | 800-811 |
| Number of pages | 12 |
| Journal | Journal of Magnetic Resonance Imaging |
| Volume | 62 |
| Issue number | 3 |
| DOIs | |
| State | Published - Sep 2025 |
Keywords
- CSE-MRI
- PDFF
- R2*
- phantom
- quantification
- reproducibility
ASJC Scopus subject areas
- Radiology Nuclear Medicine and imaging
Fingerprint
Dive into the research topics of 'Multi-Center, Multi-Vendor Validation of Simultaneous MRI-Based Proton Density Fat Fraction and R2* Mapping Using a Combined Proton Density Fat Fraction-R2* Phantom'. Together they form a unique fingerprint.Cite this
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS