TY - JOUR
T1 - Comprehensive biomechanical analysis of three reconstruction techniques following total sacrectomy
T2 - An in vitro human cadaveric model
AU - Macki, Mohamed
AU - De La Garza-Ramos, Rafael
AU - Murgatroyd, Ashley A.
AU - Mullinix, Kenneth P.
AU - Sun, Xiaolei
AU - Cunningham, Bryan W.
AU - McCutcheon, Brandon A.
AU - Bydon, Mohamad
AU - Gokaslan, Ziya L.
N1 - Funding Information:
The authors report the following disclosures (unrelated to this manuscript). Dr. Gokaslan is the recipient of research grants from DePuy Spine, AOSpine North America, Medtronic, Neurosurgery Research and Education Foundation, Integra Life Sciences, and K2M. He receives fellowship support from AOSpine North America. He holds stock in Spinal Kinetics and US Spine. Dr. Cunningham is the recipient of a grant from DePuy Spine.
Publisher Copyright:
© AANS, 2017.
PY - 2017/11
Y1 - 2017/11
N2 - OBJECTIVE: Aggressive sacral tumors often require en bloc resection and lumbopelvic reconstruction. Instrumentation failure and pseudarthrosis remain a clinical concern to be addressed. The objective in this study was to compare the biomechanical stability of 3 distinct techniques for sacral reconstruction in vitro. METHODS: In a human cadaveric model study, 8 intact human lumbopelvic specimens (L2-pelvis) were tested for flexion-extension range of motion (ROM), lateral bending, and axial rotation with a custom-designed 6-df spine simulator as well as axial compression stiffness with the MTS 858 Bionix Test System. Biomechanical testing followed this sequence: 1) intact spine; 2) sacrectomy (no testing); 3) Model 1 (L3-5 transpedicular instrumentation plus spinal rods anchored to iliac screws); 4) Model 2 (addition of transiliac rod); and 5) Model 3 (removal of transiliac rod; addition of 2 spinal rods and 2 S-2 screws). Range of motion was measured at L4-5, L5-S1/cross-link, L5-right ilium, and L5-left ilium. RESULTS: Flexion-extension ROM of the intact specimen at L4-5 (6.34° ± 2.57°) was significantly greater than in Model 1 (1.54° ± 0.94°), Model 2 (1.51° ± 1.01°), and Model 3 (0.72° ± 0.62°) (p < 0.001). Flexion-extension at both the L5- right ilium (2.95° ± 1.27°) and the L5-left ilium (2.87° ± 1.40°) for Model 3 was significantly less than the other 3 cohorts at the same level (p = 0.005 and p = 0.012, respectively). Compared with the intact condition, all 3 reconstruction groups statistically significantly decreased lateral bending ROM at all measured points. Axial rotation ROM at L4-5 for Model 1 (2.01° ± 1.39°), Model 2 (2.00° ± 1.52°), and Model 3 (1.15° ± 0.80°) was significantly lower than the intact condition (5.02° ± 2.90°) (p < 0.001). Moreover, axial rotation for the intact condition and Model 3 at L5-right ilium (2.64° ± 1.36° and 2.93° ± 1.68°, respectively) and L5-left ilium (2.58° ± 1.43° and 2.93° ± 1.71°, respectively) was significantly lower than for Model 1 and Model 2 at L5-right ilium (5.14° ± 2.48° and 4.95° ± 2.45°, respectively) (p = 0.036) and L5-left ilium (5.19° ± 2.34° and 4.99° ± 2.31°) (p = 0.022). Last, results of the axial compression testing at all measured points were not statistically different among reconstructions. CONCLUSIONS: The addition of a transverse bar in Model 2 offered no biomechanical advantage. Although the implementation of 4 iliac screws and 4 rods conferred a definitive kinematic advantage in Model 3, that model was associated with significantly restricted lumbopelvic ROM.
AB - OBJECTIVE: Aggressive sacral tumors often require en bloc resection and lumbopelvic reconstruction. Instrumentation failure and pseudarthrosis remain a clinical concern to be addressed. The objective in this study was to compare the biomechanical stability of 3 distinct techniques for sacral reconstruction in vitro. METHODS: In a human cadaveric model study, 8 intact human lumbopelvic specimens (L2-pelvis) were tested for flexion-extension range of motion (ROM), lateral bending, and axial rotation with a custom-designed 6-df spine simulator as well as axial compression stiffness with the MTS 858 Bionix Test System. Biomechanical testing followed this sequence: 1) intact spine; 2) sacrectomy (no testing); 3) Model 1 (L3-5 transpedicular instrumentation plus spinal rods anchored to iliac screws); 4) Model 2 (addition of transiliac rod); and 5) Model 3 (removal of transiliac rod; addition of 2 spinal rods and 2 S-2 screws). Range of motion was measured at L4-5, L5-S1/cross-link, L5-right ilium, and L5-left ilium. RESULTS: Flexion-extension ROM of the intact specimen at L4-5 (6.34° ± 2.57°) was significantly greater than in Model 1 (1.54° ± 0.94°), Model 2 (1.51° ± 1.01°), and Model 3 (0.72° ± 0.62°) (p < 0.001). Flexion-extension at both the L5- right ilium (2.95° ± 1.27°) and the L5-left ilium (2.87° ± 1.40°) for Model 3 was significantly less than the other 3 cohorts at the same level (p = 0.005 and p = 0.012, respectively). Compared with the intact condition, all 3 reconstruction groups statistically significantly decreased lateral bending ROM at all measured points. Axial rotation ROM at L4-5 for Model 1 (2.01° ± 1.39°), Model 2 (2.00° ± 1.52°), and Model 3 (1.15° ± 0.80°) was significantly lower than the intact condition (5.02° ± 2.90°) (p < 0.001). Moreover, axial rotation for the intact condition and Model 3 at L5-right ilium (2.64° ± 1.36° and 2.93° ± 1.68°, respectively) and L5-left ilium (2.58° ± 1.43° and 2.93° ± 1.71°, respectively) was significantly lower than for Model 1 and Model 2 at L5-right ilium (5.14° ± 2.48° and 4.95° ± 2.45°, respectively) (p = 0.036) and L5-left ilium (5.19° ± 2.34° and 4.99° ± 2.31°) (p = 0.022). Last, results of the axial compression testing at all measured points were not statistically different among reconstructions. CONCLUSIONS: The addition of a transverse bar in Model 2 offered no biomechanical advantage. Although the implementation of 4 iliac screws and 4 rods conferred a definitive kinematic advantage in Model 3, that model was associated with significantly restricted lumbopelvic ROM.
KW - Biomechanics
KW - Galveston L-rod
KW - Lumbopelvic
KW - Sacral reconstruction
KW - Sacrectomy
KW - Sacrum
UR - http://www.scopus.com/inward/record.url?scp=85032693551&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85032693551&partnerID=8YFLogxK
U2 - 10.3171/2017.2.SPINE161128
DO - 10.3171/2017.2.SPINE161128
M3 - Article
C2 - 28777063
AN - SCOPUS:85032693551
SN - 1547-5654
VL - 27
SP - 570
EP - 577
JO - Journal of Neurosurgery: Spine
JF - Journal of Neurosurgery: Spine
IS - 5
ER -