TY - JOUR
T1 - Comparison of electrochemiluminescence and ELISA methods in the detection of blood borne pathogens in Gabon
AU - Bisseye, Cyrille
AU - Ndong, Jophrette Mireille Ntsame
AU - Matoumba, Anicet Mouity
AU - Bengone, Calixte
AU - Migolet, Guy Mouelet
AU - Nagalo, Bolni Marius
N1 - Publisher Copyright:
© 2017 Hainan Medical University
PY - 2017/9
Y1 - 2017/9
N2 - Objective To assess the performances of Cobas 6000 e601 and EVOLIS BioRad in the detection of HIV, HBV and HCV in blood donors in Libreville (Gabon). Methods A cross-sectional investigation was conducted in July 2017 in a total of 2 000 blood donors recruited at the National Blood transfusion Center, Libreville Gabon. Among them, 363 donors were selected to compare the performances of COBAS 6000 e601 (electro-chemiluminescence) and EVOLIS BioRad in detecting HIV, HBV and HCV using Cohen's kappa coefficient. Results Both methods yielded similar results for the detection of HIV and HBsAg. A very good agreement of 93.39% and an excellent agreement of 98.90% were obtained for the detection of HIV and HbsAg, with kappa values of 0.80 and 0.98, respectively. The observed agreement of 91.86% was found for the detection of HCV, which gave a fair agreement between the two methods with kappa = 0.33. Conclusions The two evaluation methods showed a similar performance in the detection of HIV, HBV. However, given the high rate of intra and inter-genotypes recombination known for HIV and HBV, more robust techniques of detection such as polymerase chain reaction should be used to prevent post-transfusion contaminations.
AB - Objective To assess the performances of Cobas 6000 e601 and EVOLIS BioRad in the detection of HIV, HBV and HCV in blood donors in Libreville (Gabon). Methods A cross-sectional investigation was conducted in July 2017 in a total of 2 000 blood donors recruited at the National Blood transfusion Center, Libreville Gabon. Among them, 363 donors were selected to compare the performances of COBAS 6000 e601 (electro-chemiluminescence) and EVOLIS BioRad in detecting HIV, HBV and HCV using Cohen's kappa coefficient. Results Both methods yielded similar results for the detection of HIV and HBsAg. A very good agreement of 93.39% and an excellent agreement of 98.90% were obtained for the detection of HIV and HbsAg, with kappa values of 0.80 and 0.98, respectively. The observed agreement of 91.86% was found for the detection of HCV, which gave a fair agreement between the two methods with kappa = 0.33. Conclusions The two evaluation methods showed a similar performance in the detection of HIV, HBV. However, given the high rate of intra and inter-genotypes recombination known for HIV and HBV, more robust techniques of detection such as polymerase chain reaction should be used to prevent post-transfusion contaminations.
KW - Blood donors
KW - Cobas 6000 e601
KW - EVOLIS BioRad
KW - Kappa coefficient
UR - http://www.scopus.com/inward/record.url?scp=85028578005&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85028578005&partnerID=8YFLogxK
U2 - 10.1016/j.apjtb.2017.08.008
DO - 10.1016/j.apjtb.2017.08.008
M3 - Article
AN - SCOPUS:85028578005
SN - 2221-1691
VL - 7
SP - 805
EP - 808
JO - Asian Pacific Journal of Tropical Biomedicine
JF - Asian Pacific Journal of Tropical Biomedicine
IS - 9
ER -