Abstract
Introduction: Automated hand-held nerve conduction study (NCS) devices are being marketed for use in the diagnosis of lumbosacral radiculopathy (LSR). In this study we compared the specificity and sensitivity of a hand-held NCS device for the detection of LSR with standard electrodiagnostic study (EDX). Methods: Fifty patients referred to a tertiary referral electromyography (EMG) laboratory for testing of predominantly unilateral leg symptoms (weakness, sensory complaints, and/or pain) were included in the investigation. Twenty-five normal 'control' subjects were later recruited to calculate the specificity of the automated protocol. All patients underwent standard EDX and automated testing. Results: Raw NCS data were comparable for both techniques; however, computer-generated interpretations delivered by the automated device showed high sensitivity with low specificity (i.e., many false positives) in both symptomatic patients and normal controls. Conclusions: The automated device accurately recorded raw data, but the interpretations provided were overly sensitive and lacked the specificity necessary for a screening or diagnostic examination.
Original language | English (US) |
---|---|
Pages (from-to) | 9-13 |
Number of pages | 5 |
Journal | Muscle and Nerve |
Volume | 43 |
Issue number | 1 |
DOIs | |
State | Published - Jan 1 2011 |
Keywords
- Automated nerve conduction studies
- Electromyography (EMG)
- Lumbosacral
- Radiculopathy
ASJC Scopus subject areas
- Physiology
- Clinical Neurology
- Cellular and Molecular Neuroscience
- Physiology (medical)